Proceedings of the 2015 4th International Conference on Sensors, Measurement and Intelligent Materials

Detection and analysis of 18 PAHs in water by micro-LLE combined with concurrent solvent recondensation-large volume injection-GC-MS

Authors
Xiaolin Zhou, Bo Xu, Liya Cao, Xinying Zhang
Corresponding Author
Xiaolin Zhou
Available Online January 2016.
DOI
10.2991/icsmim-15.2016.138How to use a DOI?
Keywords
Detection, 18 PAHs, micro-LLE, GC-MS, recondensation-large volume injection
Abstract

A rapid and efficient method termed micro liquid-liquid extraction combined with concurrent solvent recondensation-large volume injection-GC-MS has been developed for the extraction and determination of PAHs in water. Some condition parameters were optimized to obtain high detection sensitivity. Calibration curves of 18 PAHs were linear in the range of 0.1-10 ng•L 1, with the correlation coefficients (r2) between 0.9993 and 1.0. The method attained good precisions from 0.3% to 4.5% for PAHs standard solutions. The method recoveries ranged between 64.5-119.5% for spiked with 5.0 ng•L 1 and 76.2-119.5% for spiked with 25.0 ng•L 1 and 50.0 ng•L 1 for real water samples. Limits of detection (LODs, S/N=3) of the method were determined from 0.010-0.068 ng•L 1. The optimized method was successfully applied to the determination of 18 PAHs in real water samples.

Copyright
© 2016, the Authors. Published by Atlantis Press.
Open Access
This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).

Download article (PDF)

Volume Title
Proceedings of the 2015 4th International Conference on Sensors, Measurement and Intelligent Materials
Series
Advances in Computer Science Research
Publication Date
January 2016
ISBN
10.2991/icsmim-15.2016.138
ISSN
2352-538X
DOI
10.2991/icsmim-15.2016.138How to use a DOI?
Copyright
© 2016, the Authors. Published by Atlantis Press.
Open Access
This is an open access article distributed under the CC BY-NC license (http://creativecommons.org/licenses/by-nc/4.0/).

Cite this article

TY  - CONF
AU  - Xiaolin Zhou
AU  - Bo Xu
AU  - Liya Cao
AU  - Xinying Zhang
PY  - 2016/01
DA  - 2016/01
TI  - Detection and analysis of 18 PAHs in water by micro-LLE combined with concurrent solvent recondensation-large volume injection-GC-MS
BT  - Proceedings of the 2015 4th International Conference on Sensors, Measurement and Intelligent Materials
PB  - Atlantis Press
SP  - 736
EP  - 739
SN  - 2352-538X
UR  - https://doi.org/10.2991/icsmim-15.2016.138
DO  - 10.2991/icsmim-15.2016.138
ID  - Zhou2016/01
ER  -